  • Seametrics TempHion™ Water Quality Sensor with Datalogging



  • Easy-to-use in-field calibration utilities

  • 316 stainless steel, fluoropolymer, acetal, and PTFE construction

  • Solution ground for excellent noise protection

  • Operates on low power

  • Long-term sensor stability

  • Measures pH, ORP, and temperature

  • RS485 network

  • 200,000+ records stored in non-volatile memory

  • Dual protocol (Modbus® and SDI-12)

  • Narrow 0.75″ diameter, 17.625″ length

  • Available cableless or with a variety of cable options—polyethylene, polyurethane, or FEP

  • Easy export to spreadsheets and databases



  • The Seametrics TempHion™ Smart Sensor is a microprocessor-based submersible sensor with built-in data logging. The internal processor in the TempHion allows for easy calibration, using the calibration utilities in Aqua4Plus. Once calibrated, this calibration data is stored in non-volatile memory within the Smart Sensor. When data is collected, this calibration information is applied to the data, resulting in highly accurate readings at a wide range of temperatures.
    The temptation is powered internally with two replaceable AA batteries. Alternately it can be powered with an external auxiliary power supply for data-intensive applications. The unit is programmed using our easy-to-use Aqua4Plus control software. Once programmed the unit will measure and collect data on a variety of time intervals.
    Several TempHions, or a combination of TempHions and other Seametrics Smart Sensors, can be networked together and controlled directly from a single computer. While most will use the TempHion with our free, easy-to-use Aqua4Plus software, it is by no means limited to that software. You can use your own Modbus® RTU or SDI-12 software or logging equipment to read measurements, thus tying into your existing telemetry and control systems.
    Another Article:

    Application

    • Single- or multi-well tracer tests
    • salt water intrusion tracking
    • Tidal influence studies
    • Wastewater treatment discharge
    • pH monitoring

    Specification

    Housing & CableWeight0.8 lb. (0.4 kg)
    Body MaterialAcetal & 316 stainless or titanium
    Wire Seal MaterialFluoropolymer and PTFE
    CableSubmersible: polyurethane, polyethylene, or ETFE (4 lb./100 ft., 1.8 kg/30 m)
    Desiccant1-3 mm indicating silica gel
    Field ConnectorStandard
    TemperatureOperating Range0 ̊ to 55 ̊C (32 ̊ to 131 ̊F)
    Storage RangeWithout batteries: -40 ̊ to 80 ̊C (-40 ̊ to 176 ̊F)
    PowerInternal BatteryTwo replaceable lithium ‘AA’ batteries – Battery life: 12 months at 15 min. polling interval (may vary do to environmental factors)
    Auxiliary12 Vdc – Nominal, 9-15 Vdc – range
    CommunicationModbus®RS485 Modbus® RTU, output=32bit IEEE floating point
    SDI-12SDI-12 (ver. 1.3) – ASCII
    LoggingMemory4MB – 200,000 records
    Logging TypesVariable, user-defined, profiled
    Logging Rates2x/sec maximum, no minimum
    Baud Rates9600, 19200, 38400
    SoftwareComplimentary Aqua4Plus 2.0
    Networking32 available addresses per junction (Address range: 1 to 255)
    File Formats.a4d and .csv (also .xls in Windows 8 and earlier)
    Output ChannelsTemperature
    pHORPBromide
    Element30K ohm thermistor, Epoxy bead/external housingGlass combination electrodePlatinum ringAg/AgCl solid-state electrode (Ion electrode method)
    Accuracy±0.2 ̊C±0.2 pH units, 0.1% mV value (typical)0.1 mVH, 0.1% mV value (typical)±5.0% of measured value (typical)
    Resolution0.1 ̊C0.01 pH units0.01 mVH units0.1 ppm
    UnitsCelsius, Fahrenheit, KelvinpH, mVEh, mVppm, mV
    Range0 ̊ to 55 ̊C (32 ̊ to 131 ̊F)0-14 pH units / -538 to 260 mV± 1200 mV0–10,000 ppm
    Compensated0 ̊ to 40 ̊C (32 ̊ to 104 ̊F)0 ̊ to 40 ̊C (32 ̊ to 104 ̊F)Isopotential point characterization
    CalibrationOne or two-point calibration w/ pH buffers (4 & 7 or 7 & 10)EH 1 pt. calibrationOne or two-point calibration w/ ionic strength adjustment
    Reference SolutionPotassium Nitrate – (KNO3)Potassium Nitrate – (KNO3)Potassium Nitrate – (KNO3)
    ReferenceAg/AgCl solid-state electrode, capillary liquid junction, TempHion™ reference solution
    Maximum DepthpH/ORP: 700 ft (210 m) / 300 psi, ISE: 230 ft (70 m)/100 psi
    EnvironmentalIP68, NEMA 6P
    *Specifications subject to change. Please consult our web site for the most current data (seametrics.com). Modbus is a registered trademark of Schneider Electric.