Cari

  • Small Vortex Flow Meter For Liquids kofloc

    Kofloc Small Karman Vortex Flow Meter For Liquids

    Kofloc FM0101/0102/0103/0105 SERIES

    KOFLOC’s Karman Vortex Flow Meter FM Series provides an ideal tool for measuring and monitoring liquid flows, including cooling water and cleaning water. Since PPS resin is used for the body material, all models of the series offer superior reliability and durability.

    Features

    • Simple design that minimizes a dead space
    • Measurements of very small flows available (up to 0.5 L/min)
    • Use of PPS resin has achieved a small, lightweight and rigid
    • The sensor can be used for pure water or deionized water or deionized water and chemicals (the sensor is acid-/alkali-resistant).
    • These models are in the process of application for CE Marking

    Standard Specifications

    Model
    FM0101
    FM0102
    FM0103
    FM0105
    Dimension
    (a)
    17.8
    17.5
    17.5
    32.5
    Dimension
    (L)
    80.6
    80
    80
    110
    Connection
    (X)
    R3/8
    R1/2
    R1/2
    25A
    Flow
    range
    0.5-4 L/min
    2-16 L/min
    4-40 L/min
    10-150 L/min
    Fluids
    for measurement
    Cooling
    water, cleaning water, etc.
    Measuring
    accuracy
    Within
    ±3.0% F.S
    Repeatability
    Within
    ±0.5% F.S
    Outputs
    S Type:
    4-20mA
    P-Type:
    Pulse (Open collector) (For w/o indicator only)
    D Type:
    With indicator
    Supply
    voltage
    12-24VDC
    Liquid
    temperature range
    0-70 Celcius
    Proof
    pressure
    1MPa
    Amb.
    temperature range
    0 – 50 Celcius
    Amb.
    humidity range
    5-90%RH
    Applicable
    cleanliness/
    IP64
    (Splashproof construction per JIS C 0920)
    waterproofing
    standards
    Material
    for wetted part
    PPS
    with 30% glass mixture
    PPS w/o
    glass mixture
    Cable
    length
    W/o
    indicator: 2 meters long; terminated/pretended (pre-soldered)
    With
    indicator: 3 meters long; terminated/pretended (pre-soldered)
    Weight
    W/o
    indicator: 85 g (Sensor unit)
    165 g
    (Sensor unit)
    With
    indicator: 100 g (Sensor unit)
    205 g
    (Sensor unit)

    Principle Measurement

    When a columnar object (an object that generates vortices) is placed in the flow path a of fluid, regulated Karman vortex channels, is generated at the back of the object.
    Since the frequency of a vortex generated is linearly proportional to the flow velocity within a given range, the flow amount can be measured by counting the number of vortices.
    These series models make use of the principle. When the frequency of each vortex generated is detected by an incorporated vortex detector (piezoelectric device), the signal processing circuit outputs a signal which is linearly proportional to volume flow.
  • Entri Populer