Cari

  • Kofloc DF-550C SERIES All-Metal Digital Mass Flow Controller

    DF-550C is now released as an all-metal model of the high-performance digital mass flow controller DF-250C Series. The metal sealed structure enables controlling the measurement of corrosive gases such as semiconductor process gases.

    Features

    • Response improved to 1 sec or faster in all flow rate ranges thanks to a new solenoid
    • The stainless steel seal supports high-reactive gases.
    • Lineup ranging from 10 SCCM to 10 SLM
    • Two communication ports enabling daisy-chaining
    • RS-485 communication function installed as standard
    • RoHS CE standard compliant
      Another Article:

    Specification

    Sensor typeThermal sensor
    Valve typeNormally closed proportional solenoid valve
    Applicable gases*1N2, Air, H2, He, Ar, O2, CO2, etc.
    Gases not specifically mentioned here are subject to calibration for N2 equivalent.)
    Flow range*210 SCCM to 10 SLM
    Control range2 to 100% (F.S.)
    Response*3Total flow rate control range ≤ 1 sec (within ±2% F.S.)
    Accuracy*4±1.0%F.S.
    Repeatability±0.25%F.S.
    Operating differential pressure*550 to 350 kPa (100 to 350 kPa for F.S. 10 SLM)
    Inlet maximum pressure500kPa(G)
    Proof pressure980kPa(G)
    External leak rate≤ 1.0 x 10-11 Pa·m3/sec (He)
    TemperatureWorking temp.5 to 50℃
    Accuracy guaranteed temp.15 to 35℃
    Allowable storage temp.-10 to 60℃
    Allowable operating humidity10 to 90% RH (without dew condensation)
    Materials of gas contact partSUS316L, PTFE, and magnetic stainless steel
    Electrical connectionD-sub 9-pin, KFC standard (SEMI standard compliant), RJ-45 modular jacks (two)
    Flow rate setting signal0 to 5 V DC (input impedance approx. 1 MΩ)
    Flow rate output signal0 to 5 V DC (load resistance 10 KΩ)
    Digital communicationAddress setting with RS485 rotary switch: up to 99 devices (9600 bps)
    Required power supply (DC)+15 V DC (±5%), 150 mA; -15 V DC (±5%), 150 mA
    Joint1/4 VCR or equivalent
    Mounting postureHorizontal installation recommended.
    WeightApprox. 1100 g
    *1 The gases must be dry and clean, free of corrosive components and foreign matter such as dust and mist.
    *2 The flow rate units SCCM and SLM indicate a mass flow rate converted to a volume flow rate in cc/min and L/min at 0°C and 1 atm
    *3 Guarantee for calibration gas: N2.
    *4 Accuracy for calibration gas: N2, flow range (max. full scale).
    *5 The operating differential pressure may differ depending on specifications.
    For further information please contact : marketing@wmablog.com